The following is an example of typical output obtained on entry of the print Parameters command. The information in lower case is explanatory and is not output.
LWL 20 UPL 85 setting of detector
set lower level at 20%, upper rate at 85%.
RFS 2E3 full scale setting of ratemeter 2E3.
RTC .5 RSP 10 (RTC S) recorder time constant -5 second, recorder
speed 10 mm/degree.
SPE 0.02 INT 1.00 scan speed 0.02 deg., sampling interval time
1 seconds.
STP 0.050 MXN 10 step size 0.05degrees, number of
measurement points on either side of angle.
MXS 0.010 MXT 2.00 step of 0.01 degrees between measuring
points. measurement 1 second.
BGT 5.00 BGM 2 background measurement 5 seconds. backround 2.
BGA 0.700 WAV 1.54184 background offset 0.7 degrees. wavelength
1.54184.
SOF ROF DON POF spinner off. recorder/ratemeter OFF. dual
output on. printing recording off.
ZERO2T -4.990 set the goniometer "ZERO"-position.
10/25/2004 JQ (originally SS then GL 01/01/98)